Welcome to Guangdong Fleming Electronic Technology Co.,Ltd. !
Back

UES48DZ-SPA

Product Features:
Meet edical safety approvals 
2 MOPP input to output isolation
Leakage current≤100μA 
Up to 5,000m operating altitude 
Energy  efficiency level Vl
≤0.1W standby power

Input:   
Input Voltage Range 90-264VAC
Frequency Range 47-63Hz 
Input Current  1.3A at 90VAC 
Inrush Current 100A max at 240VAC cold start 
Touch Leakage Current(Max)  < 100μA at 264VAC

Environmental:   
Operating Temperature -0ºC to 40ºC 
Storage Temperature -40ºC to 80ºC 
Operating Humidity 10% to 90% RH, non-condensing 
Storage Humidity 5% to 95% RH 
Operating Altitude 5,000m 

Protection:   
Overload 110-160% rated output power, auto recovery 
Over Voltage Min120% rated output voltage input to reset 
Short Circuit Trip and restart (hiccup mode) 

Safety Approvals:     
Safety aAgency/Mark Medical
CB  IEC60601-1
TUV-SUD/mark   EN60601-1 

Mechanical Details     
110.0(L)*51.5(W)*33.1(H)mm   
We are a professional medical power supply wholesaler. These products are power supplies for medical devices. They are sold directly from the factory with low prices, fast delivery, and have safety certifications from various countries.