Welcome to Guangdong Fleming Electronic Technology Co.,Ltd. !
Back

UES12LCP-SPA-OP

• Meets medical & I.T.E. safety

• 2 MOPP input to output isolation

• Touch current ≤ 10μA

• Energy efficiency level VI

• ≤ 0.075W standby power

• 4V to 24V outputs, up to 12W

• Up to 5,000m operating altitude


Model NumberVoltage (V)Current (A)Rated PowerRipple & Noise (max)Voltage toleranceLine& Load RegulationEfficiency (Average)Start Up delay

UES12LCP-XXXYYYSPA-OP

UES12LCP-XXXYYYSPC-OP

4.0-5.00.01-2.0010.00W150±7%Line:±1%
Load:±5%
79.01%≤3s
5.1-6.00.01-2.0012.00W150±7%80.19%≤3s
6.1-7.00.01-1.7111.97W150±7%83.30%≤3s
7.1-8.00.01-1.5012.00W150±7%83.30%≤3s
8.1-9.00.01-1.3311.97W150±5%83.30%≤3s
9.1-10.00.01-1.2012.00W150±5%83.30%≤3s
10.1-11.00.01-1.0911.99W150±5%83.30%≤3s
11.1-12.00.01-1.0012.00W150±5%83.30%≤3s
12.1-13.00.01-0.9412.22W200±5%83.30%≤3s
13.1-14.00.01-0.8612.04W200±5%83.30%≤3s
14.1-15.00.01-0.8012.00W200±5%83.30%≤3s
15.1-16.00.01-0.7512.00W200±5%83.30%≤3s
16.1-17.00.01-0.7112.07W200±5%83.30%≤3s
17.1-18.00.01-0.6712.06W200±5%83.30%≤3s
18.1-19.00.01-0.6311.97W200±5%83.30%≤3s
19.1-20.00.01-0.6012.00W200±5%83.30%≤3s
20.1-21.00.01-0.5711.97W200±5%83.30%≤3s
21.1-22.00.01-0.5512.10W200±5%83.30%≤3s
22.1-23.00.01-0.5211.96W200±5%83.30%≤3s
23.1-24.00.01-0.5012.00W200±5%83.30%≤3s

1.jpg

Notes

(*1) Other options are available, please contact our sales representative for details.

(*2) Measured at output connector with 20MHz bandwidth and 0.1uF ceramic in parallel with 10uF electrolytic capacitors.

Input Voltage Range

90-264VAC
Frequency Range 47-63Hz
Input Current0.5A at 90VAC
Inrush Current50A max at 240VAC cold start
Touch Leakage Current≤10μA at 264VAC


Dimensions(mm)

71.5(L)*42(W)*19(H)
Weight35g
MTBF>100,000hrs MIL-HDBK-217 at 25℃


Overload120-200% rated output power, auto recovery
Over Voltage120-150% rated output voltage input to reset
Short CircuitTrip and restart (hiccup mode)


Safety Agency / MarkMedicalITE
CB

IEC60601-1

IEC60950-1    
ULANSI/AAMI ES60601-1-
CAN/CSA C22.2 NO. 60601-1
FCC-FCC PART15


EmissonsMedicalITE
ConductedIEC/EN 60601-1-2,CISPR 11EN55032, CISPR 32
RadiatedIEC/EN 60601-1-2,CISPR 11EN55032, CISPR 32
Harmonic CurrentsEN61000-3-2, Class AEN61000-3-2, Class A
Voltage FlickerEN61000-3-3EN61000-3-3
ImmunityIEC/EN 60601-1-2EN55024, CISPR 24
Radiated ImmunityEN61000-4-3

10V/m ,3V/m 80MHz - 2.7GMHz

EFT/BurstEN61000-4-4±2KV on AC port, ±1KV on signal ports
SurgeEN61000-4-5±2KV line to line (different mode) compatible with ±4KV line to line (different mode) 
Conducted ImmunityEN61000-4-63Vrms, 6Vrms (0.15MHz-80MHz)
Magnetic FieldEN61000-4-830 A/m

Dips & Interruptions

EN61000-4-110%, 70%, 0% of UT


Dielectric Withstand Voltage

5,656VDC input to output

Insulation Resistance10M Ohms, 500VDC input to output